Dentsply Sirona Nitram DAC Universal S Cleaning, Lubrication & Sterilization Unit
Couldn't load pickup availability
Dentsply Sirona Nitram DAC Universal S — Lubrication, Sterilization & Cleaning
DAC UNIVERSAL helps reduce hygiene risks, protect your team and patients from cross-contamination, and support legal compliance — including best practice conditions referenced in Health Technical Memorandum 01-05 and the UK Department of Health (2009) priority for cleanliness in dental settings (as required in other countries).
Standards Compliance
- Compatible with European standard EN 13060 Class S small sterilizers for cleaning turbines, micromotors, and contra-angles
- Cleaning efficacy compliant with ISO 15883:5 Annex J
Process Control
Document the full program cycle with the DAC UNIVERSAL printer and digitally. Recorded data include selected program, date, time, batch number, actual vacuum, pressure and temperature values, confirmation of successful completion, and any unauthorized deviation.
One-button hygiene: close the special lid and start the fully automatic cleaning program. Instruments are cleaned, lubricated, and sterilized — reducing operator error and supporting consistent results.
Cycle Steps
- Internal cleaning with cold water
- Fully automatic lubrication
- External cleaning with hot and cold water
- Sterilization and drying
Equipment Options
- Unwrapped dynamic instruments: Load up to six dynamic handpieces on dedicated adapters for cleaning, maintenance, and sterilization; fit adapters to the handpieces used in your clinic
- Wrapped dynamic instruments: Clip up to three packaged handpieces individually with spring clips; instruments first run an unpackaged cycle, then are packaged before sterilization
- Solid instruments: Place up to five solid instruments (e.g., burs, mirrors, curettes) in the supplied wire basket
Cleaning Time Comparison (4–6 devices)
- DAC UNIVERSAL: approx. 16 min (semi-critical B); approx. 35 min* for critical B ready for immediate use
- Manual cleaning: approx. 20 min / approx. 60 min
- Cleaning and disinfection device with subsequent manual lubrication: approx. 45 min / approx. 1 h 10 min
- Semi-automatic cleaning and maintenance device with subsequent sterilization: approx. 60 min / approx. 1 h 35 min
- * Wrapped instrument sterilization (standard in Class B autoclaves)
Sterilization Programs
- Standard program 134°C, 3 min sterilization — total cycle approx. 16 min
- Standard prion program 134°C, 18 min sterilization — total cycle approx. 32 min
- Standard maintenance program 121°C, 15 min sterilization — total cycle approx. 28 min
- Program for wrapped instruments* 134°C, approx. 3 min sterilization — total cycle approx. 22 min
- * Lid required for wrapped products; cycle times may vary with the operating environment
Economic & Care Benefits
- Cost per handpiece can be two to three times lower than other automatic cleaning methods; consumption cost four to five times lower than manual preparation methods
- Lubrication system tested and approved by international manufacturers for micromotors, contra-angles, and turbines
- Automatic preparation, fast return to use, lower capital and maintenance/approval costs, discounted repair pricing
- Optional Sirona NitraDem Direct Connect water treatment for fully automatic water supply; DAC PROFESSIONAL can drain used water automatically when connected to the drain system